1. Product Usage
The use of xenon arc lamps that can simulate the full sunlight spectrum to reproduce destructive light waves in different environments can provide corresponding environmental simulation and accelerated test basis for scientific research, product development and quality control.
Aging tests are conducted by exposing material samples to the light and heat radiation of xenon arc lamps to evaluate the light resistance and weather resistance of certain materials under the action of high-temperature light sources. It is mainly used in paints, coatings, rubber, plastics, pigments, adhesives, fabrics, cosmetics, aerospace, ships and vessels, automobiles, electronics, packaging industries, etc.
2. Standards
- - GB/T16422.2
- - GB/T1865-97
- - GB/T9344-88
- - ASTMG155
- - GB/T2423.24-1995
- - ISO10SB02/B04
- - SAEJ2527
- - SAEJ2412
- - DIN 75220
3. Technical specifitions
Model |
TL-SF-80 |
TL-SF-270 |
TL-SF-500 |
TL-SF-1000 |
Inner box size mm (W*D*H) |
700×400×300 |
750×600×600 |
800×800×800 |
1000×1000×1000 |
Outer box size mm (W*D*H) |
900×900×1900 |
1320×1050×1830 |
1400×1250×2030 |
1600×1450×2230 |
Temperature range |
RT+10℃~70℃ |
|||
Blackboard temperature |
40 ℃~100℃ adjustable |
|||
Humidity range |
30% ~98%RH adjustable |
|||
Temperature uniformity |
≤±3℃ |
|||
Temperature accuracy |
≤±0.1℃ |
|||
Humidity accuracy |
≤±1%RH |
|||
Lamp source |
3 air-cooled lamps |
|||
Lamp brand |
Teelong / Q-LAB (optional) |
|||
Lamp life |
About 1200 hours |
|||
Lamp power |
2.5KW domestic; 1.8KW imported |
|||
Distance from arc center to specimen |
200 ~400mm Effective exposure area 2400cm2 |
|||
Photoperiod time |
0 ~ 999H, M, S continuously adjustable |
|||
Rainfall time |
0 ~ 9999min continuous rainfall adjustable |
|||
Rainfall cycle |
1 ~ 240min interval (break) rainfall adjustable |
|||
Water spray cycle |
18min/102min or 12min/48min (water spraying time/no water spraying time) |
|||
Rain water pressure |
0.12~0.15Mpa |
|||
Water nozzle aperture |
Ф0.8mm |
|||
Distance between sample holder and lamp |
200 ~ 500mm adjustable |
|||
Filter Type |
Simulate light through glass window, solar filter, UV filter, you can choose one |
|||
Irradiation intensity |
Full spectrum 290~2450nm, 550~1120 W/m2 |
|||
Or the irradiance of ultraviolet 340nm band is 0.35~1.2W/m2 |
||||
Or the irradiance of ultraviolet light in the 300-400NM band is 40W-60W |
||||
Or the irradiance of ultraviolet 420NM band is 0.55~1.5W/m2 |
||||
Equipment use conditions |
Voltage: AC380V±10%; Frequency: 50±0.5Hz; Power: 16KW three-phase five-wire system |
|||
Environmental conditions |
Temperature 5℃~30℃; relative temperature ≤85%RH; air pressure 86kPa~106kPa |
|||
Safety protection device |
Leakage, short circuit, over-temperature, water shortage, motor overheating, over-current protection |